Immunohistochemistry Staining Machine

Categoría:
Specification:
Model Number : HS-2500
Short Desc : Fully Automated Immunohistochemical Stainer
The fully automatic immunohistochemical staining instrument is based on the basic principle of immunology – antigen-antibody reaction, that is, the principle of specific binding of antigen and antibody. It uses chemical reaction to make the color developer (fluorescein, enzyme, metal ion, isotope) of the labeled antibody develop color to determine the antigen (peptide and protein) in the tissue cells, and conduct localization, qualitative and relative quantitative research on it. This product is used for sample processing before pathological analysis (dewaxing, antigen repair, staining)

Features:
1. The same instrument automatically completes baking, dewaxing, pretreatment, immunohistochemical staining, silver staining in situ hybridization and in situ hybridization operations. Various types of slides can be tested at the same time and support multiple staining schemes;
2. Processing method: single-frame batch processing, four-frame independent operation;
3. Function: dewaxing, repair, cleaning, primary/secondary antibody dripping, DAB ready-to-use, reagent refrigeration, temperature control;
4. Companion diagnosis: ALK, PDL-1, her-2 and other companion diagnosis projects can be carried out;
5. Throughput: 48 slides (12*4);
6. Speed: 48 slides in less than 3.5 hours (Leica process);
7. Sample type: tissue/enriched cell specimens;
8. Sample addition method: support manual addition of antibody reagents, easy to observe and adjust;

9. Reagent position: 44 11*4;
10. Reagent bottle capacity: 15ml;
11. Barcode management: reagent 1D code, specimen 2D code;
12. Liquid transfer method: steel needle, air exchange, capacitive sensing liquid level, closed counting;
13. System liquid: non-toxic waste liquid (10L), toxic waste liquid, cleaning liquid, pure water, repair liquid, xylene, 100% ethanol, 30% hydrogen peroxide;
14. Waste liquid treatment: waste liquid can be automatically discharged, no need to treat harmful waste liquid;
15. Repair function: citrate buffer or other repair liquid, 90-1 00℃ heated with liquid; *Repair liquid requires special treatment;

16. Heating and cooling time: from 25℃ to 95℃, no more than 10min
from 95℃ to 30℃, no more than 15min
17. Temperature control accuracy: 95℃+-5℃;
18. Reagent cooling: 8-15℃, fluctuation +-2℃;
19. Reaction incubation: 30-50℃ adjustable, fluctuation +-2℃;
20. Liquid coverage: special siphon;
21. Cleaning method: special siphon;
22. Liquid transfer accuracy: 30ul C.V≤5%;
23. Liquid transfer range: 30ul-500ul 10ul increments;
24. The amount of primary antibody added is controlled at 150ul;
25. Special functions: whipping and mixing, reagents are used as they are prepared, self-checking on and off, and the experimental process can be edited;
26. Interface: CAN communication;
27. Optional external devices: computer, printer;
28. Customized cover glass and reagent bottles are supplied as consumables;
29. Through the hydrophilic diffusion of the cover plate, it can be automatically and evenly covered.;
30. During the equipment test, each glass slide is automatically covered with a liquid sealing film to ensure the integrity of the sample and the non-evaporation of the reagent, and the tissue incubation environment is stable;
31. It has data storage function and quality control system;
32. Scanning system: with barcode scanning, fully automatic identification of samples and reagents;
33. Instrument weight: about 135kg
34. Instrument size (length, width and height): 1010mm x 760mm x 810mm

Packing Information:
Weight : 270kgsDimension(mm) : 1150*920*1020mm
710*520*530mm
800*800*700mm
Packing : In wooden case